CALIBRATION CHAPTER 5
Press Unit (F2) key and select the desired torque unit.
10. Using the Up () and Down () Cursor keys, enter the Full Scale value for the 4-in-1 transducer to calibrate.
Possible values are:
1 = 50 in. lb.
2 = 400 in. lb.
3 = 1000 in. lb.
4 = 250 ft. lb.
( For 4 in 1 Transducers display shows : CHECKING TRANSDUCER.)
Apply torque to one of the transducers until “ Full Scale UNITS shows on display.
Note: Example graphics are for 50 in-lb transducer.
11. Press ENTER ().
12. Press any key to initiate Zerotare. The LCD displays:
then:
13. Apply torque using the precision arm, hanger or certified weights listed as calibration point 1 in the
calibration check point table in this chapter. When the reading is stable, push and hold the Up () or Down
() Cursor Keys to get the correct torque value reading. Repeat the procedure for points 2, 3, 4, and 5 of the
same table as shown below:
Apply 10% Torque = 5 in.lb
5-9